Product name
|
Glass Ionization Vacuum Gauge
|
|
Model
|
model GI-D7
|
|
Measurable pressure range
|
WIT:1.30×10-5 to 6.7×10-1 Pa
WIB:1.30×10-6 to 2.3×10-2 Pa
|
|
Accuracy
|
±15%
|
|
Pressure indication
|
Didital indication 2 digit mantissa, 1 digit exponent
|
|
Compatible sensor head
|
WIT-G1, WIT-G2, WIT-G3, WIB-G5, WIB-N3
|
|
Sampling time
|
100ms
|
|
DEGAS
|
Direct electrical heating of grid AC power 1.5VA
|
|
Functions
|
Range hold function, Gas sensitivity setting, Filament selection
|
|
Measurement value output
|
linear output 0 to 10 VDC, Pseudo-logarithmic output 0-10VDC(1V linear signal in each digit), Log output
|
|
Control input
|
REMOTE/LOCAL, Filament ON/OFF, Filament 1/2, DEGAS ON/OFF, Range hold
|
|
Control output
|
Analog output(0-10V), BCD output, Filament ON/OFF, Emission current setting, Degassing ON/OFF
|
|
Set point(contact point capacity)
|
2 point(AC125Vmax, 0.5Amax / DC24Vmax, 1.0Amax)
|
|
Serial communication spec
|
RS232C
|
|
Operating temperature range
|
10 to 40℃(50 to 104℉)
|
|
Operating humidity range
|
15 to 18%(not condensing)
|
|
Power supply voltage
|
AC100±10V 50/60Hz, 120VA
|
|
Fuse
|
BS. SEMKO 3.15A (2pcs)
|
|
External dimensions W×D×H
|
240mm×380mm×99mm
|
|
Weight
|
5.7kg
|
|
Standard accessories
|
Power cord
|
3P Plug with ground
|
|
Connector
|
For external input / output
|
|
Quick manual
|
1 pc
|
|
Options
|
Sensor cable
|
4, 10, 15, 20, 25, 30m
|
|
Gauge head
|
Triode type glass gauge:WIT-G1, WIT-G2, WIT-G3
B-A type glass gauge:WIB-G5
B-A tupe nude gauge:WIB-N3
|